| Reported | Class | Firm | Product and reason |
|---|---|---|---|
| Sep 23 | Class II | Wilson-Cook Medical Inc. Device |
Brand Name: HEMOSPRAY ENDOSCOPIC HEMOSTAT Product Name: HEMOSPRAY ENDOSCOPIC HEMOSTAT Model/Catalog Number: HEMO-7, HEMO-10, HEMO-7-EU, HEMO-10-EU Software Version: N/A Product Description: Hemostatic device for endo… Due to issue with endoscope adherence and inability to spray powder. Firm will provide an updated Instruction for Use (IFU). |
| Apr 22 | Class II | Wilson-Cook Medical Inc. Device |
Instinct Plus Endoscopic Clipping Device REF: G58010 Rx Only, Sterile EO Due to increased in complaints their is the potential for endoscopic clipping device to malfunction. |
| Feb 25 | Class II | Wilson-Cook Medical Inc. Device |
Teslatome Bipolar Sphincterotomes REF: TESLA-B2535-260 Model Number: G60784 UDI-DI code: 00827002607842 REF: TESLA-B2535-450 Model Number: G60785 UDI-DI code: 00827002607859 These devices are used for cannulat… Due to product that may contain a void in the material that separates the wire guide and cutting wire lumens, which may cause an inappropriate path of current during cutting. |
| Feb 11 | Class II | Wilson-Cook Medical Inc. Device |
Product Name: Nasal Jejunal Feeding Tube REF: NJFT-10 Nasal feeding tube packaged without the nasal transfer tube component. |